New TopShim run:

version = topshim 2.2.0
date = 2024-06-04
time = 20:47:15.225
command = topshim 

FINE SHIMMING

Parameters:

probe = Z163739_0737
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 3.5 Hz, Target= 0.4 Hz, After= 1.0 Hz
Iteration 2: FWHM Before= 1.0 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration 3: FWHM Before= 0.4 Hz, Target= 0.4 Hz, After= 0.5 Hz
Iteration 4: FWHM Before= 0.5 Hz, Target= 0.6 Hz, After= 0.8 Hz
Iteration 5: FWHM Before= 0.8 Hz, Target= 0.5 Hz, After= 0.5 Hz
Iteration 6: FWHM Before= 0.5 Hz, Target= 0.3 Hz, After= 0.7 Hz
Iteration 7: FWHM Before= 0.7 Hz, Target= 0.5 Hz, After= 0.6 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D shimming, non-spinning) = 0.782

shim changes: 
Z	-480
Z2	-112
Z3	-197
Z4	+180
Z5	+79
Z6	-1827
Z7	-4544
Z8	-1574

duration = 3 min 17 sec
completed successfully
finished Tue Jun  4 20:50:29 2024


========================================================================